An End cap for an Intra- medullary nail is normally used to prevent bone overgrowth into the proximal end of the nail and to facilitate nail removal at a later date. End cap migration into the proximal redundant space in the medullary canal behind the nail is rare but an avoidable complication. Maintenance of a fixed position of knee flexion during reaming and nail insertion and the use of Cannulated End Cap systems with Image control and may reduce the incidence of this complication.
For the retrieval of the end cap from the canal a haemostatic forceps may be used.
How to cite : Goel A, Maheshwari M , Niyazi M S, Gupta A, End cap missing the proximal end of the tibial interlocking nail to migrate posteriorly into the medullary canal: A complication that may be avoided. Indian J Orthop Surg 2019;5(1):99-100
This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.