Get Permission Joshi and Patil: Combination of exchange nailing augmented with plate and allograft in treating femoral shaft nonunion after intramedullary nailing – Our experience


Introduction

Femur is the largest bone in the body. Fractures of the femoral shaft often are the result of high energy trauma and may be associated with multiple injuries.1 Main goal of treatment is restoration of length, alignment, rotation, and preservation of blood supply and avoiding infection.2 There are various modalities for management of femur shaft fractures such as closed reduction and cast immobilization, skeletal traction, intramedullary nailing, interlocking intramedullary nailing (antegrade or retrograde) and plating. Locked intramedullary nailing is currently considered to be the treatment of choice for most femoral shaft fractures.3 Intramedullary nailing is preferred as plating has higher risk of infection and nonunion.4 Risk factors for femoral nonunion after intramedullary nailing are open fracture, fracture in distal third of femur treated with antegrade nailing, delay in weight bearing, and tobacco use.5 The treatment of nonunion with a statically locked intramedullary nail in place is either bone grafting in situ, dynamization of the nail by locking bolt removal, or exchange nailing.6 We combined exchange nailing augmented with plate and allografting to study its usefulness in treating femur shaft nonunion after nailing.

Table 1

The patient profiles are shown below

S.No.

Age

Sex

Site

Non-Union Time [month]

Solid union time [wk.]

Operation Time [min]

Intraoperative blood loss[ml]

Remark

1

30

M

M/3

18

24

120

190

Kuntscher nail

2

28

F

M/3

22

18

150

250

IM locking nail

3

40

M

L/3

25

22

180

500

IM locking nail

4

33

M

U/3

20

21

125

350

Broken screw

5

25

M

M/3

18

18

122

300

IM locking nail

6

60

M

L/3

19

27

130

250

Distal dynamized

7

32

M

U/3

20

20

140

400

IM locking nail

8

28

M

U/3

22

18

130

300

IM locking nail

9

35

M

M/3

15

18

120

280

Distal dynamized

10

56

F

M/3

16

19

90

250

IM locking nail

11

39

M

M/3

22

15

130

350

Distal dynamized

12

36

F

M/3

23

22

140

300

IM locking nail

13

29

M

L/3

20

18

120

450

IM locking nail

14

33

F

L/3

21

20

100

340

IM locking nail

15

37

F

L/3

20

21

130

320

Failed to unite.

Materials and Methods

We did a retrospective study. Cases were collected between Aug 2020-June 2022. 15 patients were collected, 10 men and 5 women. They had undergone intramedullary nailing for femur fracture. Fractures which went into nonunion after intramedullary nailing were typically either juxta-articular fractures i.e., either subtrochanteric fracture of femur or fractures in distal third of femur treated with antegrade femoral nailing with only 2 interlocking bolts in distal fragment. Infection was ruled out. Mean period for presentation of nonunion - 20 months after primary surgery. All the patients had uninfected nonunion. 10 patients had undergone intramedullary nailing with proximal and distal bolts locked, 3 had distal dynamized intramedullary nailing, 1 had breakage of proximal bolt of intramedullary nail and 1 had Kuntscher nail. Initially the old nail was removed. Reaming was done. Then we freshened the ends of the nonunion till the ends bled. Then we passed a nail, did distal interlocking with three bolts (two transverse and one anteroposterior) and then added a plate. Plate used was 4.5mm system 8–10-hole DCP contoured for femur. We used a muller device over the plate to compress the nonunion and added an allograft when needed. Minimum 3 screws in plate on either side of fracture were aimed. The nail was finally locked proximally.

Figure 1

A, B, C): X-ray AP, LAT, OBLIQUE views on Initial presentation of patient with femur non-union with nail in situ with cerclage wires 18 months after primary surgery; C): Fracture site opened with freshening of edges; D): Mullers device over plate used to achieve compression at fracture site; E – C): Arm image showing compression achieved over fracture site with mullers device; F): After application of derotation plate and Allografts at nonunion; G): Immediate postoperative x-ray; H): X-ray showing full union after 18 months in all views

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/ba693fa5-b0a7-4014-9c91-9586603703e6-uimage.png

Another patient with femur non-union

Figure 2

Preoperative x-ray

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/d330a5d5-4cc1-4cdb-9246-98af99dd46d7-uimage.png

Figure 3

Immediate postop

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/80007ebc-fb1b-492d-8f41-48ca74c7e2ef-uimage.png

Figure 4

18 month follow-up x-ray

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/cc87425e-5333-4f77-8c81-7e4f1e0e6c72-uimage.png

Figure 5

Clinical pictures of patient showing range of motion and scar

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/c7b26223-42f5-41b7-9197-084040655c1b-uimage.png

Figure 6

One patient presented with subtrochanteric fracture which was treated with intramedullary nail and went into nonunion

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/2643cf64-3ed7-406e-a45a-9e9ea551f865-uimage.png

Figure 7

Muller device used for compression

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/06f0b768-a19e-45f2-8066-7e163cb9442d/image/947c5145-313e-4c5a-b0cd-242e44735b92-uimage.png

Results

Table 2

Distribution depending on age

Age in years

Frequency

Percentage

<30

4

26.6

30-50

9

60

>50

2

13.3

Total

15

100

Table 3

Distribution depending on sex

Gender

Frequency

Percentage

Male

10

66.66

Female

5

33.33

Total

15

100

Table 4

Distribution depending on site of nonunion

Site of femur Nonunion

Frequency

Percentage

Upper 1/3

3

20

Middle 1/3

7

46.6

Lower 1/3

5

33.3

Total

15

100

Table 5

Distribution depending on non-union time

Non-union time in months

Frequency

Percentage

10-19

5

33.33

20-30

10

66.66

Total

15

100

All but one patient achieved postoperative bony union uneventfully at a mean time of 22.1weeks (range, 12–40 weeks). The mean operative time was 105 minutes (range, 60–150minutes), and the mean blood loss was 300 ml (range, 150–500 ml). Immediate weight bearing started as tolerated by patient. The 14 patients could walk bearing full weight without pain within 3 months. There were no significant complications such as broken hardware, implant back-out, axial or rotational malalignment, or deep infections in the successful united patient. One patient did not achieve union after the procedure.

Discussion

Achieving union of long bone diaphyseal fractures, away from isthmus of bone, where the medullary cavity is wide, remains a challenge.7 This is especially true when treated with intramedullary nailing. Because nailing is a load sharing device and provides only relative stability.8 Therefore, not adequate rotational stability is present where medullary canal is wide. On the other hand, plating does provide rigid fixation avoiding rotational instability.9 With plating, chances of implant breakage remain an issue as it is a load bearing implant.10 Both nail and plate alone are inadequate in treating nonunion of juxta articular long bone fractures. Combining plate which provides torsional stability with nail which provides axial and bending stability provides best chance of treating these nonunion. It is wwell-establishedthat to treat femur non unions one can go for exchange nailing augmented with plating. We combined the above with allograft to ensure union and avoid patient the pain of repeat surgery preventing additional morbidities associated with surgery. We used allograft when there was gap at fracture after nibbling of sclerosed edges of nonunion, and used local hypertrophic cancellous bone when gap was minimal. One should be generous while removing the sclerosed ends as failure to do so would prevent union despite of reaming exchange nailing augmented plating and allografting. We learnt this the hard way which led to one failure to unite in our study. Using Judet’s technique (1963), osteoperiosteal flaps were raised. This elevation of cortical chips that remains attached to the periosteum surrounding the non-union site physiologically helped in increasing the decorticated surface area (which acts as multiple fractures and induces bone formation), exposure of vascular subcortical Haversian system, Osteogenic stimulation by inflammation of mechanical origin and acts as ‘in situ’ vascularized autogenous bone graft. Other similar studies were done by ekre-echam, where they used exchange nailing and augmented plating as a combined procedure in two cases and achieved union in both; Wang et al. where they used thick interlocking nail augmented with plate fixed with unicortical screws only; and Sancheti et al. where they used exchange K-nailing augmented with plating fixed with bicortical screws and autograft. But our study differs from these in that we used interlocking nail and a mullers device to approximate the fracture ends and an allograft.

Conclusion

Exchange nailing with plate and graft augmentation is a good option in treating femur shaftnonunion after open reduction and internal fixation with intramedullary nailing. We used exchange nailing augmented with plate and bone graft and found it to lead to union in 93.3% cases.

Source of Funding

None.

Conflict of Interest

None.

References

1 

UE Anyaehie OC Ejimofor FC Akpuaka CU Nwadinigwe Pattern of femoral fractures and associated injuries in a Nigerian tertiary trauma centreNiger J Clin Pract20151844626

2 

MS Jones B Waterson Principles of management of long bone fractures and fracture healingSurgery (Oxford)202038291910.1016/j.mpsur.2019.12.010

3 

HK Pihlajamäki ST Salminen OM Böstman The treatment of nonunions following intramedullary nailing of femoral shaft fracturesJ Orthop Trauma2002166394402

4 

JE Shroeder R Mosheiff A Khoury M Liebergall YA Weil The outcome of closed, intramedullary exchange nailing with reamed insertion in the treatment of femoral shaft nonunionsJ Orthop Trauma20092396537

5 

LA Taitsman JR Lynch J Agel DP Barei SE Nork Risk factors for femoral nonunion after femoral shaft fractureJ Trauma2009676138992

6 

EA Swanson EC Garrard DT Bernstein DP OʼConnor MR Brinker Results of a systematic approach to exchange nailing for the treatment of aseptic femoral nonunionsJ Orthop Trauma2015291217

7 

TE Mcmillan AJ Johnstone Technical considerations to avoid delayed and non-unionInjury201748Suppl 1648

8 

S Weller D Höntsch Medullary nailing of femur and tibiaManual of internal fixationSpringerBerlin, Heidelberg1992291365

9 

SE Tan ZJ Balogh Indications and limitations of locked platingInjury200940768391

10 

C Sommer R Babst M Müller B Hanson Locking compression plate loosening and plate breakage: a report of four casesJ Orthop Trauma20041885717



jats-html.xsl


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.