Get Permission Vyas, Batajoo, and Misra: Pull out suture using the plunger flange of the syringe, a better alternative to the button


Introduction

Tendon avulsions within zone 1 of the hand represent a complex and challenging subset of injuries that require careful consideration and innovative surgical approaches. Traditionally, the button technique, first introduced by Bunnell1 in 1944, has been a cornerstone for repairing tendon avulsions by anchoring the sutured tendon to the bone using a button secured over the finger nail. This method, though pioneering, has undergone modifications to enhance its applicability. For instance, Mantero et al2 further adapted the technique by situating the button over the fingertip, eliminating the need to pass it through the distal phalanx. The intricate nature of zone injuries necessitates a systematic classification for effective management. Moimen and Elliot's3 classification divides laceration injuries within zone 1 into three distinct subgroups (Figure 1). Group 1a encompasses lacerations distal to the A5 pulley, presenting a challenge in placing core sutures due to the anatomical complexity. Zone 1b injuries reside between the distal edge of the A4 pulley and zone 1a, while zone 1c encompasses injuries beneath the A4 pulley. Consequently, zones 1b and 1c often lend themselves to conventional tendon repair techniques, while zone 1a avulsions pose a unique dilemma.4

Figure 1

Sub-classification of zone 1 FDP injury

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/8e3ebb24-1eae-4fa6-a682-9652a4484613/image/1338699d-ba64-4a6e-bd3d-b1d7c31b515a-uimage.png

The management of avulsions or complete disruptions of the flexor digitorum profundus tendon within zone 1a commonly involves the application of a pullout suture technique tied over a button, as advocated by Bunnell.1 This technique, while established, has been scrutinized for its limitations, prompting the exploration of innovative alternatives to optimize tendon repair outcomes.5

The current study investigates a unique alternative to the traditional button technique by utilizing the plunger flange of a syringe. This novel approach presents the potential to address the challenges associated with the button technique, particularly in terms of procedural efficiency, availability, and early postoperative rehabilitation.

While the button technique has demonstrated its utility, it is not without limitations. Challenges in locating appropriately sized buttons in the fast-paced environment of an operating theater have raised concerns regarding procedural delays and the potential impact on patient outcomes. In contrast, the plunger flange technique offers the advantage of simplicity, immediate availability within operation theaters, and a potentially smoother surgical workflow.

To explore the feasibility and effectiveness of the plunger flange technique, this study compares its outcomes with those of the traditional button approach in patients with zone 1a and 1b injuries. By systematically evaluating aspects such as surgical feasibility, postoperative complications, and the timeline for initiating therapy, this study aims to contribute valuable insights to the field of tendon avulsion repair within zone 1 of the hand.

The overarching goal of this study is to introduce and validate the plunger flange technique as a potentially superior alternative to the traditional button method, thereby expanding the armamentarium of surgical options available for managing tendon avulsions within zone 1 of the hand.

Materials and Methods

Technique

In case of zone 1a and 1b injuries the routine method is reinsertion of tendon onto distal phalanx using a modification of the technique as described by Bunnell. Here a two-strand 3-0 or 4-0 Polypropylene core tendon suture is passed through the bone and nail complex and tied over a button. However, in this study we have used plunger flange instead of button.

Numerous techniques have been developed to repair flexor digitorum profundus tendon after avulsion or laceration.5 Commonly internal or external fixation using bone anchor or buttons are preferred.

In our study we have used plunger flange instead of button.

To substitute the button; firstly, we have removed the plunger flange from the syringe and then drilled 2 small holes on the plunger flange with a 1.5 mm k-wire (Figure 2). A 3-0 polypropylene suture is then passed through these holes, after that we have placed a piece of sterile gauze between the 2 threads and tied with a simple knot (Figure 3) resembling button system.

Figure 2

a): Holes are being made into separated plunger end via K-wire and drill machine; b): Separated plunger end is shown with 2 holes made in it; c): Enlarged view of plunger end with holes, for passage of suture; d): Sutures passed via modified plunger end after extensor tendon repair

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/8e3ebb24-1eae-4fa6-a682-9652a4484613/image/e5a45fea-9b9b-430a-bc17-c3a3a719e071-uimage.png

Figure 3

A sterile gauze between 2 threads and tied with a simple knot over the plunger flange

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/8e3ebb24-1eae-4fa6-a682-9652a4484613/image/ce97092a-1510-4835-9b19-a823ecf3e082-uimage.png

We also diverge the holes towards pulp in mallet fracture and try to avoid nail if applying from volar side.

Or larger syringe can be used depending on the size necessary (Figure 4, Figure 5)

This technique allows patients to start therapy soon after surgery. The first dressing is done after 10 days; however, proximal joints can be mobilized during that time. At 6 weeks we remove plunger flange in office or OPD after which involved joint can be mobilized.

Figure 4

a): A 10 ml syringe; b): A 5 ml and 10 ml syringes with plunger pulled out and showed separately

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/8e3ebb24-1eae-4fa6-a682-9652a4484613/image/bd416864-e1e5-4cd9-91d2-b4e0a2671ea6-uimage.png

Figure 5

a, b): Both sides of the pushing end of plunger shown separately

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/8e3ebb24-1eae-4fa6-a682-9652a4484613/image/fbd16a0a-a910-4511-9aaa-2d258a4e39d1-uimage.png

Results

The plunger flange technique demonstrated notable feasibility and efficacy in the repair of tendon avulsions within zone 1, thereby shedding light on its potential as a valuable alternative to the conventional button method.

The study underscores the value of the plunger flange technique as a viable alternative for repairing tendon avulsions within zone 1. While acknowledging the shared complications, the technique's enhanced availability, ease of preparation, and procedural efficiency present compelling attributes that warrant further exploration. The outcomes of this study contribute to the evolving discourse on innovative methods for improving the management of tendon injuries within this challenging anatomical region.

Discussion

In the realm of tendon repair techniques,6, 7 the plunger flange emerges as an intriguing alternative to the conventional button approach, offering a range of advantages that address longstanding challenges. One of the prominent advantages lies in the simplicity of procurement and preparation of the plunger flange, a departure from the logistical complexities of finding appropriately sized buttons. Particularly in bustling operation theaters where time-sensitive decisions are commonplace, the ease of access to plunger flanges could significantly streamline surgical procedures.

The enhanced availability of plunger flanges in every operation theater provides a noteworthy benefit. Unlike buttons, which might require specific sizing and planning, plunger flanges are ubiquitous due to their standard presence in various medical contexts. This inherent availability aligns well with the dynamic and unpredictable nature of clinical settings, offering surgeons a readily accessible tool to facilitate efficient and consistent repair procedures. The plunger flange's accessibility is in stark contrast to the scenario where button availability can lead to delays and compromises in patient care.

In comparing the usability of plunger flanges to the traditional button technique, it's important to consider the convenience factor. Plunger flanges can be swiftly prepared, and their use is straightforward, reducing the intricacies associated with the traditional button technique. This convenience might have a cascading effect, potentially expediting the overall surgery and allowing medical professionals to focus more on the surgical nuances rather than grappling with technicalities. Furthermore, the relative simplicity of the plunger flange technique might facilitate its adoption among surgeons, ultimately leading to a broader and more consistent application.

However, as with any medical procedure, a balanced consideration of potential complications is imperative. While the plunger flange technique exhibits advantages, its associated complications must be acknowledged. Interestingly, the complications associated with the plunger flange technique closely resemble those of the traditional button method.8, 9, 10 These complications encompass issues such as local wound irritation, nail deformities, skin necrosis, pain, infection, and the potential for snagging or rupture of the repair.11, 12, 13 The overlap in complications between the two techniques underscores the importance of rigorous postoperative monitoring and care,14, 15 regardless of the method employed.

It is noteworthy that a comprehensive understanding of these complications can inform surgeons in anticipating and mitigating potential risks, ensuring that patients receive optimal care throughout their recovery journey. Further, studies may be warranted to elucidate any nuances or variations in complication rates between the plunger flange technique and the traditional button approach, enabling a more nuanced assessment of the risk-benefit profile.

Conclusion

Our technique using the planger flange is simple and easily reproducible in operation theatre. Contrary to button, the plunger flange doesn’t need to be sterilized separately; it is also readily available across an array of sizes. Thus, the authors recommend the plunger flange as a better alternative to the button.

Source of Funding

None.

Conflict of Interest

None.

References

1 

S Bunnell Surgery of the HandJ.B. Lippincott Company United States1944

2 

R Mantero P Bertolotti C Badoino Il pull-out in “no man’s land” e al canale digitale nelle lesioni dei flessori (metodo personale)Riv Chir Mano19731111930

3 

NS Moiemen D Elliot Primary flexor tendon repair in zone 1J Hand Surg Br20002517884

4 

S Huq S George DE Boyce Zone 1 flexor tendon injuries: a review of the current treatment options for acute injuriesJ Plast Reconstr Aesthet Surg2013668102331

5 

JP Leddy JW Packer Avulsion of the profundus tendon insertion in athletesJ Hand Surg Am197721669

6 

CA Goldfarb RH Gelberman MI Boyer Flexor tendon reconstruction: current concepts and techniquesJ Am Soc Surg Hand20055212330

7 

R Singh B Rymer P Theobald PB Thomas A review of current concepts in flexor tendon repair: physiology, biomechanics, surgical technique and rehabilitationOrthop Rev (Pavia)201574612510.4081/or.2015.6125

8 

N Kang D Marsh D Dewar The morbidity of the button-over-nail technique for zone 1 flexor tendon repairs. Should we still be using this techniqueJ Hand Surg Eur Vol200833556670

9 

ASC Bidwai L Feldberg The button-over-nail technique for zone I flexor tendon injuriesHand Surg201217033659

10 

D Guinard F Montanier D Thomas D Corcella F Moutet The Mantero flexor tendon repair in zone 1J Hand Surg Br199924214851

11 

MB Geary KK Li RC Chadderdon RG Gaston Complications following transosseous repair of zone I flexor tendon injuriesJ Hand Surg Am202045121183

12 

K Tsuge Y Ikuta Y Matsuishi Repair of flexor tendons by intratendinous tendon sutureJ Hand Surg Am19772643640

13 

E Smock N Kang Should we still be using buttons for Zone 1 flexor tendon repairsPlast Reconstr Surg2019144472526

14 

JB Tang Outcomes and evaluation of flexor tendon repairHand Clin20132922519

15 

NM Cannon JW Strickland Therapy following flexor tendon surgeryHand Clinics19851114765



jats-html.xsl


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.